![Flight Medical Innovations 60 DL Скачать руководство пользователя страница 47](http://html.mh-extra.com/html/flight-medical-innovations/60-dl/60-dl_service-manual_2305321047.webp)
Removing and Reinstalling Modules
Replacing the O2 mixer Neb board * (iO2 Internal mixer model only)
Flight 60 Service Manual
|
4-17
12.
Disconnect the O
2
sensor cable.
13.
Remove the front panel from the ventilator.
4.8.2 Connecting the Front Panel Assembly
To connect the front panel assembly:
Follow the above procedure in reverse order.
Replacing the O
2
mixer Neb board
*
(
iO
2
Internal mixer
model only
)
4.9.1 Removing the O
2
mixer neb board
To remove the O
2
mixer neb board:
1.
Follow the procedures for removing the front panel assembly (see section 4.8 )
2.
remove the 3 screws securing the O
2
mixer Neb board to the Main Board.
Figure 34 – Securing screws
3.
Pull the mixer board out of the connector .
4.9.2 Installing the O
2
mixer board
To connect the O
2
mixer board:
Follow the above procedure in reverse order.