![Flight Medical Innovations 60 DL Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/flight-medical-innovations/60-dl/60-dl_service-manual_2305321056.webp)
Removing and Reinstalling Modules
Replacing the Manifold Assembly
4-26
| FLIGHT 60 Service Manual
Figure 50 – Removing the Manifold Assembly
The vibration dumping frame is an integral part of the manifold and should not
be disassembled from the manifold.
Before sending the manifold assembly to FLIGHT MEDICAL, remove the short
and long flex tubing.
4.12.2 Installing the Manifold Assembly
To install a manifold assembly:
Perform the removal procedure in the reverse order, with the following
additional step after inserting the muffler assembly, which is helpful for getting
the cable ties through their slots:
Insert the detachable battery until it is almost completely in, and then assemble
the cable ties through the slots under the muffler.
If a new manifold was installed, run PCS downloader to update the manifold
serial number and reset working hours. See section 5.2.