Removing and Reinstalling Modules
Replacing the O2 mixer filters * (iO2 Internal mixer model only)
Flight 60 Service Manual
|
4-31
Replacing the O
2
mixer filters
*
(
iO
2
Internal mixer
model only
)
To replace the filters:
1. Use twizers to remove the O-ring that secures the filters
2. Use the twizers to pull out the filters (2 each)
3. Place the new filters (2 each) at the O
2
input and push all the way in
4. Push the O-ring in, verify it's in the groove, securing the filters in place
Figure 56 – O2 regulator filter
Replacing the Purge Board Assembly
4.17.1 Removing the Purge Board Assembly
To remove the purge board assembly:
1. Remove the detachable battery (see section 4.3) and the internal battery (see
section 4.4).
2. Remove the ventilator cover (see section 4.5).
3. Remove the front panel (see section 4.8).
4. Disconnect the two check valves at the ends of the silicon tubes from the purge
board.