Software Upgrade
Calibration integrity check
Flight 60 Service Manual
|
6-5
Vti
Within the 450-550 spec.
Pbase 5
±
1cm H
2
O
O
2
72% - 88% for External O
2
Mixer.
76% - 84% for Internal O
2
mixer (iO
2
) model
5.
If any of those values is out of spec, perform the appropriate calibration and
check again.