Proline t-mass 65 MODBUS RS485
GAS
Hauser
39
13
GAS
Function description, GAS group
Prerequisite
This function is
not
available if the IN-SITU CALIBRATION function is enabled. This function is available again if the
in-situ calibration is reset to the factory setting. Contact your Hauser service organization for more
information.
Description
Use this function to view or change the gas configuration
• The device can be configured with one or two independent gas groups.
• Switching between two groups can be performed manually (function SELECT GROUP
40)
General programming rules:
• A gas group can be configured with 1 single gas or a gas mixture (max. 8 gas constituents).
• The sum of the gas mixture constituents entered must total 100.0 Mole %.
• A gas constituent and its corresponding Mole % can be entered in any order within the mixture.
• A gas constituent may have a value of 0.0 Mole % within the mixture.
• The option NOT USED is only a place holder for functions GAS TYPE 1 and GAS TYPE 2 to 8 where there is no gas
assigned. The program does not use this option for calculations.
• The option SPECIAL GAS is a substitute for other gases. The Mole % value is always 100.0%
!
Note!
For a flowchart of the GAS QUICK SETUP, see Operating Instructions BA00115D/06
Programming examples
a.
1 gas group: 1 standard gas
GAS GROUP 1
1
AIR
100.0 %
b.
2 gas groups: 2 standard gases
GAS GROUP 1
GAS GROUP 2
1
1
ARGON
NITROGEN
100.0 %
100.0 %
c.
2 gas groups: 1 standard gas, 1 special gas
GAS GROUP 1
GAS GROUP 2
1
1
OXYGEN
SPECIAL GAS
100.0 %
100.0 %
–
O2 90% OZONE 10%
–
1.2009
–
1.5005 kg/m3