TOTALIZER 1/2
Proline t-mass 65 MODBUS RS485
28
Hauser
8
TOTALIZER 1/2
Function description, TOTALIZER 1/2 group
ASSIGN TOTALIZER
Description
Use this function to assign a measured variable to the totalizer.
• The totalizer is reset to "0" as soon as the option selected is changed.
• If you select OFF, only the ASSIGN TOTALIZER function is displayed in the
Totalizer 1 or 2 group.
Options
0 = OFF
1 = MASS FLOW
2 = CORRECTED VOLUME FLOW
3 = HEAT FLOW
Factory setting
HEAT FLOW
MODBUS register:
TOTALIZER 1
TOTALIZER 2
Data type:
Access:
2601
2801
Integer
read/write
SUM
Description
The total for the totalizers measured variable aggregated since measuring
commenced appears on the display. The value is positive.
The totalizers' response to errors is defined in the FAILSAFE MODE function
(
30).
Display:
Max. 7-digit floating-point number, including unit (e.g. 15467.04 kg)
MODBUS register:
TOTALIZER 1
TOTALIZER 2
Data type:
Access:
2610
2810
Float
read
OVERFLOW
Description
The total for the totalizers overflow aggregated since measuring commenced
appears on the display.
Total flow is represented by a floating-point number consisting of max.
7 digits. Higher numerical values (>9,999,999) can be viewed as overflows.
The effective quantity is thus the total of the SUM function (
28) plus the
value displayed in the OVERFLOW function.
Example:
Reading after 2 overflows: 2 E7 kg (= 20000000 kg)
The value displayed in the SUM function = 196845.7 kg
Effective total quantity = 20196845.7 kg
Display:
Integer with exponent, including unit, e.g. 2 E7 kg
MODBUS register:
TOTALIZER 1
TOTALIZER 2
Data type:
Access:
2612
2812
Float
read
UNIT TOTALIZER
Description
For selecting the unit for the measured variable assigned to the totalizer.
Options (ASSIGN TOTALIZER = MASS FLOW):
SI
0 = g;1 = kg; 2 = t
US
3 = oz; 4 = lb; 5 = ton
Factory setting
kg or lb (country dependent
55)
Options (ASSIGN TOTALIZER = CORRECTED VOLUME FLOW):
SI
0 = Nl, 1 = Nm³
US
2 = Sm³, 3 = Scf
Factory setting
Nm³ or Sm³ (country dependent
55)
Options (ASSIGN TOTALIZER = HEAT FLOW):
SI
0 = kWh, 1 = MWh, 2 = kJ, 3 = MJ, 4= GJ, 5 = kcal, 6 = Mcal, 7 = Gcal
US
8 = kBtu, 9 = MBtu, 10 = GBtu, 11 = tonh
Factory setting
MWh or kBtu (country dependent
55)
MODBUS register:
TOTALIZER 1
TOTALIZER 2
Data type:
Access:
2602
2804
Integer
read/write