QUICK SETUP
Proline t-mass 65 MODBUS RS485
16
Hauser
5.3
Quick Setup "Sensor"
It is essential that the insertion sensor is setup according to the actual pipe or duct and then installed
at the calculated insertion depth. This Quick Setup guides you systematically through the procedure
to setup the sensor.
!
Note!
The QUICK SETUP SENSOR function is not available for flanged type sensors.
A0009910-en
PIPE TYPE
m
• CIRCULAR
– in case that the pipe is of a standard type, then parameterize functions PIPE STANDARD and NOMINAL
DIAMETER
– In case that the pipe is a non-standard type, then select OTHERS in the function PIPE STANDARD and
parameterize the functions WALL THICKNESS and OUTER DIAMETER.
– The function INTERNAL DIAMETER displays the calculated internal diameter and is read only.
• RECTANGULAR
– Enter the INTERNAL HEIGHT, INTERNAL WIDTH and WALL THICKNESS of the duct
– Select the MOUNTING orientation of the sensor: HORIZONTAL or VERTICAL
MOUNTING SET LENGTH
n
Enter the measured length of the mounting set (including the compression fitting).
INSERTION DEPTH
o
This function calculates the insertion depth value for the mounting of the sensor.
This function is read only.
Press
F
to save settings and return to QUICK SETUP SENSOR group.
+
+
E
Esc
E
+
-
XXX.XXX.XX
HOME-POSITION
+
E
m
n
o
INSERTION
DEPH
MOUNTING SET
LENGTH
QUICK SETUP
QS
SENSOR
PIPE
STANDARD
INTERNAL
HEIGHT
NOMINAL
DIAMETER
INTERNAL
WIDTH
UNIT
LENGTH
WALL
THICKNESS
WALL
THICKNESS
MOUNTING
RECTANGULAR
CIRCULAR
OUTER
DIAMETER
INTERNAL
DIAMETER
PIPE
TYPE
Selection Pipe Type