Proline t-mass 65 MODBUS RS485
OPERATION
Hauser
23
6
OPERATION
Function description, OPERATION group
LANGUAGE
Description
For selecting the language in which all messages are shown on the local display.
If you press the
P
keys simultaneously at startup, the language defaults to
"ENGLISH".
Options
0 = ENGLISH
1 = DEUTSCH
2 = FRANCAIS
3 = ESPANOL
4 = ITALIANO
5 = NEDERLANDS
6 = NORSK
7 = SVENSKA
8 = SUOMI
9 = PORTUGUES
10 = POLSKI
11 = CESKI
Factory setting
country dependent
MODBUS register:
Data type:
Access:
2502
Integer
read/write
ACCESS CODE
Description
All data of the measuring system are protected against inadvertent change.
Programming is disabled and the settings cannot be changed until a code is
entered. If you press the
P
operating elements in any function, the measuring
system automatically goes to this function and the prompt to enter the code
appears on the display (programming disabled).
You can enable programming by entering the private code (factory setting = 65,
see DEFINE PRIVATE CODE function (
23).
• The programming levels are disabled if you do not press the operating
elements within 60 seconds following a return to the HOME position.
• Programming can also be disabled by entering any number (other than the
private code).
• Your Hauser representative can be of assistance if you mislay your
private code.
!
Note!
This function is only relevant for the local display and does not affect data
transmission via the MODBUS communication.
User input
Max. 4-digit number: 0 to 9999
MODBUS register:
Data type:
Access:
2508
Float
read/write
DEFINE PRIVATE CODE
Description
Use this function to specify the private code for enabling programming.
• Programming is always enabled if the code defined = 0.
• Programming has to be enabled before the code can be changed. When
programming is disabled this function cannot be edited, thus preventing
others from accessing your personal code.
User input
Max. 4-digit number: 0 to 9999
Factory setting
65
MODBUS register:
Data type:
Access:
2510
Float
read/write