QUICK SETUP
Proline t-mass 65 MODBUS RS485
18
Hauser
Setting or viewing the active gas group
Go to the function SELECT GROUP (
40) and simply select 1 or 2 and then exit using
ESC (
Q
keys simultaneously). No save function is necessary.
!
Note!
This Quick Setup Gas function is not available if an in-situ calibration function has been performed
on the device as the in-situ calibration curve refers to the sensor power at each recorded flow point.
Therefore, the programmed gas settings become redundant.
Performing the Quick Setup
1.
GAS GROUP
Use the
O
or
S
key to select the required GAS GROUP and continue with
F
.
– select the NUMBER OF GASES in the group from 1 to 8
– select the GAS TYPE from the choose list.
– enter the MOLE % for each GAS TYPE (only if NUMBER OF GASES is 2 and more).
– The error message CHECK VALUES appears if the total mixture % does not equal 100%. Go
back and check the mixture settings.
2.
SAVE CHANGES?
– Select YES to save changes to the GAS GROUP and activate them. Press
F
to continue or
– Select CANCEL to save the entered settings in buffer memory but not activate them for
measurement. If this function is selected, then it will be necessary to come back to this gas
group and save it at a later stage.
– Select DISCARD to clear the last changes and return to CONFIGURE GROUP to make new
settings.
3.
ANOTHER GAS GROUP?
– Select YES to continue to the CONFIGURE GROUP function. Use the + or - key to select
another group and proceed as per above instructions.
– Select NO to exit to the Quick Setup.
!
Note!
You can find more detailed information on the GAS GROUP in chapter GAS