Proline t-mass 65 MODBUS RS485
MEASURING VALUES
Hauser
7
3
MEASURING VALUES
Function description, MEASURING VALUES group
!
Note!
The engineering unit of the measured variable displayed here is configured in the SYSTEM UNITS group (
MASS FLOW
Description
The currently measured mass flow appears on the display.
Display:
5-digit floating-point number, including unit
e.g. 462.87 kg/h; 731.63 lb/min
MODBUS register:
Data type:
Access:
2007
Float
read
CORRECTED VOLUME FLOW
Description
The calculated corrected volume flow appears on the display.
The corrected volume flow is calculated from the measured mass flow and the
reference density of the gas.
Display:
5-digit floating-point number, including unit
e.g. 104.97 Nm3/h; 110.73 Sm3/h; etc.
MODBUS register:
Data type:
Access:
2011
Float
read
HEAT FLOW
Description
The calculated heat flow appears on the display.
Display:
5-digit floating-point number, including unit,
(e.g. 175.00 kJ/h; 50.000 kBtu/h; etc.)
MODBUS register:
Data type:
Access:
2005
Float
read
TEMPERATURE
Description
The currently measured temperature appears on the display.
Display:
5-digit fixed-point number, incl. unit and sign
e.g. –23.4 °C, 160.0 °F, 295.4 K
MODBUS register:
Data type:
Access:
2017
Float
read