QUICK SETUP
Proline t-mass 65 MODBUS RS485
12
Hauser
QUICK SETUP HEAT FLOW
Description
Use this function to start the application-specific setup for heat flow.
For a flowchart of the QUICK SETUP HEAT FLOW:
Options
0 = NO
1 = YES
Factory setting
NO
T-DAT SAVE/LOAD
Description
Use this function to save the configuration/settings of the
transmitter
to a
transmitter-DAT (T-DAT), or to load a configuration from the T-DAT to the
EEPROM (
manual
backup function).
For a flowchart of the T-DAT SAVE/LOAD:
Application examples:
• After commissioning, the current measuring point parameters can be saved to
the T-DAT as a backup.
• If the transmitter is replaced for some reason, the data from the T-DAT can
be loaded into the new transmitter (EEPROM).
Options
0 = CANCEL
1 = SAVE (from the EEPROM to the T-DAT)
2 = LOAD (from the T-DAT to the EEPROM)
!
Note!
• If the target device has an older software version, the message "TRANSM.
SW-DAT" is displayed during startup. Then only the "SAVE" option is
available.
• LOAD
This option is only possible:
– if the target device has the same software version as, or a more recent
software
version than, the source device or
– if the T-DAT contains valid data that can be called up
• SAVE
This function is always available.
Factory setting
CANCEL
MODBUS register:
Data type:
Access:
2401
Integer
read/write
Function description, QUICK SETUP group