HANDLING TOTALIZER
Proline t-mass 65 MODBUS RS485
30
Hauser
9
HANDLING TOTALIZER
Function description, HANDLING TOTALIZER group
RESET ALL TOTALIZERS
Description
Resets the sums and overflows of both totalizers to "zero" (=RESET).
If the device is equipped with a status input and if it is appropriately configured,
a totalizer reset (totalizers 1 and 2) can also be triggered by a pulse (see ASSIGN
STATUS INPUT function on Page
31).
Options
0 = NO
1 = YES
Factory setting
NO
MODBUS register:
Data type:
Access:
2609
Integer
read/write
FAILSAFE MODE
Description
For selecting the behavior of the totalizer in an alarm condition.
Options
0 = STOP
The totalizer does not continue to count the flow if a fault is present. The
totalizer stops at the last value before the alarm condition occurred.
1 = ACTUAL VALUE
The totalizers continue to count on the basis of the current flow data. The fault
is ignored.
2 = HOLD VALUE
The totalizer continues to count the flow on the basis of the last valid flow data
(before the fault occurred).
Factory setting
STOP
MODBUS register:
Data type:
Access:
2607
Integer
read/write