STATUS INPUT
Proline t-mass 65 MODBUS RS485
32
Hauser
SIMULATION STATUS INPUT
Description
Use this function to simulate the status input, i.e. to trigger the function
assigned to the status input (see ASSIGN STATUS INPUT function (
31).
• Notice message #671 to 652 "SIM. STATUS INPUT" indicates that simulation
is active.
• The measuring device continues to measure while simulation is in progress
i.e. the current measured values are output correctly via the other outputs.
!
Note!
The setting is not saved if the power supply fails.
Options
0 = OFF
1 = ON
Factory setting
OFF
MODBUS register:
Data type:
Access:
4306
Integer
read/write
VALUE SIMULATION STATUS
INPUT
Prerequisite
This function is
only
available if ON was selected in the SIMULATION STATUS
INPUT function (
32).
Description
Use this function to define the level to be assumed by the status input during
simulation. This value is used to test downstream devices and the measuring
device itself.
When you exit This function and the SIMULATION STATUS INPUT function is
activated (= ON), the message "END SIMULATION?" appears on the display. If
you select YES, the simulation is terminated. If you select NO, the simulation is
continued.
!
Note!
The setting is not saved if the power supply fails.
Options
0 = LOW
1 = HIGH
Factory setting
LOW
MODBUS register:
Data type:
Access:
4307
Integer
read/write
Function description, STATUS INPUT group