Technical data
Proline Prosonic Flow 93C FOUNDATION Fieldbus
78
Hauser
Link Master (LM)
support
Yes
Link Master
Factory setting
Basic Device
Selectable
Device basic current
11 mA
Device starting current
<11 mA
Device error current (FDE)
0 mA
Device (lift off)
min. voltage
9 V (H1-segment)
Permissible fieldbus supply
voltage
9 to 32 V
Integrated reverse polarity
protection
Yes
ITK Version
5.01
Number of VCRs (total)
48
Number of link objects in VFD
40
Device capacitance
In accordance with IEC 60079-27, FISCO/FNICO
Galvanic isolation
All circuits for inputs, outputs and power supply are galvanically isolated from each other.
Data transmission speed
31.25 kbit/s, voltage mode
Signal coding
Manchester II
Bus times
Min. idle time between two telegrams:
MIN_INTER_PDU_DELAY = 6 octet time (transfer time per octet)