Proline Prosonic Flow 93C FOUNDATION Fieldbus
Troubleshooting
Hauser
65
No. # 1xx
→
Software error
111
Device status message (FF):
Totalizer could not be restored
at startup –
Err. No. 111
Local display:
S: CHECKSUM TOTAL.
$
: # 111
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Cause:
Checksum error of the totalizer.
Remedy:
1.
Restart the measuring device.
2.
Replace the amplifier board.
Transducer_Error = Electronics
failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
121
Device status message (FF):
Software compatibility
problem amplifier –
I/O module –
Err. No. 121
Local display:
S: A / C COMPATIB.
!: # 121
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Cause:
Due to different software versions, I/O board and
amplifier board are only partially compatible
(possibly restricted functionality).
!
Note!
– This is indicated on the display as a warning
message for only 30 seconds (with entry in error
history).
– This situation in which the software versions
differ can occur if only one electronics board has
been exchanged; the extended software
functionality is not available. The previously
existing software functionality is still available
and measurement operation is possible.
Transducer_Error = I/O failure
(input/output error)
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
No. # 2xx
→
Error in DAT / no communication
205
Device status message (FF):
Save to T-DAT failed –
Err. No. 205
Local display:
S: LOAD T-DAT
!: # 205
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Cause:
Data backup (download) to T-DAT failed, or error
when accessing (uploading) the calibration values
stored in the T-DAT.
Remedy:
1.
Check whether the T-DAT is correctly plugged
into the amplifier board.
2.
Replace T-DAT if faulty.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics. Check the:
– Spare part set number
– Hardware revision code
3.
Replace measuring electronics boards if
necessary.
Transducer_Error = Electronics
failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
206
Device status message (FF):
Restore from T-DAT failed –
Err. No. 206
Local display:
S: SAVE T-DAT
!: # 206
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Transducer_Error = Electronics
failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
No.
Error messages:
FOUNDATION Fieldbus
(FF)*
(local display)
Transducer Block
error messages
"Diagnosis" Transducer
Block
Analog Input Function Block
error messages
Reason for error/rectification
(spare parts
→
ä
71)