Technical data
Proline Prosonic Flow 93C FOUNDATION Fieldbus
82
Hauser
10.1.7
Operating conditions: installation
Installation instructions
Any orientation (vertical, horizontal)
Restrictions and other installation instructions
Inlet and outlet runs
Version
Length of connecting cable
Shielded cables are available in the following lengths:
5 m (16.4 ft), 10 m (32.8 ft), 15 m (49.2 ft) and 30 m (98.4 ft)
Route the cables well clear of electrical machines and switching elements.
10.1.8
Operating conditions: environment
Ambient temperature range
• Prosonic Flow 93 transmitter:
–20 to +60 °C (–4 to +140 °F)
• Measuring tube Prosonic Flow C with measuring sensors Prosonic Flow W:
–10 to +80 °C (+14 to +176 °F)
• PVC sensor cable:
–20 to +70 °C (–4 to +158 °F)
!
Note!
• Install the transmitter in a shady location.
Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
• In the case of heated pipes, or pipes with cold fluids, it is permitted to completely insulate the
measuring tube with the mounted ultrasonic sensors.
Storage temperature
The storage temperature corresponds to the ambient temperature range of the transmitter, the
measuring tube, the sensors and the corresponding sensor cable
Degree of protection
• Prosonic Flow 93C transmitter: IP 67 (NEMA 4X)
• Prosonic Flow W sensors: IP 68 (NEMA 6P)
Shock resistance
Following IEC 68–2–6
Electromagnetic compatibility
(EMC)
In accordance with IEC/EN 61326 and NAMUR Recommendation NE 21
10.1.9
Operating conditions: process
Medium temperature range
Measuring tube Prosonic Flow C with measuring sensors Prosonic Flow W:
–10 to +80 °C (+14 to +176 °F)
!
Note!
Drinking Water Approval: 0 to +60 °C (+32 to +140 °F)
Limiting medium pressure
range (nominal pressure)
Perfect measurement requires that the static fluid pressure is higher than vapor pressure. The
maximum nominal pressure is PN 16 (16 bar / 232 psi).
Pressure loss
There is no pressure loss.
Pressure loss occurs if adapters are used upstream and downstream of the device. The corresponding
values can be taken from the chart