Proline Prosonic Flow 93C FOUNDATION Fieldbus
Troubleshooting
Hauser
67
592
Device status message (FF):
Channel initialization run CH1
– Err. No. 592
Local display:
S: INIT. RUN CH1
$
: # 592
BLOCK_ERR = No Error
OUT. QUALITY = UNCERTAIN
Cause:
Initialization running.
Remedy:
Wait until process is finished.
OUT. SUBSTATUS = Non specific
593
Device status message (FF):
Channel initialization run CH2
– Err. No. 593
Local display:
S: INIT. RUN CH2
$
: # 593
BLOCK_ERR = No Error
OUT. QUALITY = UNCERTAIN
OUT. SUBSTATUS = Non specific
No. # 6xx
→
Simulation mode active
602
Device status message (FF):
Positive zero return active CH1
– Err. No. 602
Local display:
S: POS. 0-RET. CH1
!: # 602
BLOCK_ERR = No Error
OUT. QUALITY = UNCERTAIN
Cause:
Positive zero return is active.
Remedy:
Deactivate positive zero return:
• FF: Transducer Block "Transducer_CH1"
(base index: 1400) or Transducer Block
"Transducer_CH2" (base index: 1500)
→
Parameter "Sys. - Positive Zero Return"
→
OFF
• Local display: BASIC FUNCTIONS
→
SYSTEM PARAMETERS
→
CONFIGURATION
→
POS. ZERO RETURN (
→
OFF)
OUT. SUBSTATUS = Non specific
603
Device status message (FF):
Positive zero return active CH2
– Err. No. 603
Local display:
S: POS. 0-RET. CH2
!: # 603
BLOCK_ERR = No Error
OUT. QUALITY = UNCERTAIN
OUT. SUBSTATUS = Non specific
604
Device status message (FF):
Positive zero return active
CH1&2 – Err. No. 604
Local display:
S: POS. 0-RET. CH1&2
!: # 604
BLOCK_ERR = No Error
OUT. QUALITY = UNCERTAIN
OUT. SUBSTATUS = Non specific
691
Device status message (FF):
Simulation failsafe active –
Err. No. 691
Local display:
S: SIM. FAILSAFE.
!: # 691
BLOCK_ERR = Simulation
active
OUT. QUALITY = UNCERTAIN
Cause:
Simulation of response to error (outputs) is active
Remedy:
Switch off simulation:
• FF: "Diagnosis" Transducer Block (base index:
1600)
→
"Sys. – Sim.Failsafe Mode" parameter
→
OFF
• Local display: SUPERVISION
→
SYSTEM
→
OPERATION
→
SIM. FAILSAFE
MODE (
→
OFF)
Transducer_Error = No Error
OUT. SUBSTATUS = Non specific
692
Device status message (FF):
Simulation Volume flow active
- Err. No. 692
Local display:
S: SIM. MEASURAND
!: # 692
BLOCK_ERR = Simulation
active
OUT. QUALITY = UNCERTAIN
Cause:
Simulation of measured variable is active.
Remedy:
Switch off simulation:
• FF: "Flow" Transducer Block (base index: 1400)
→
"Simulation - Measurand" parameter
→
OFF
• Local display: SUPERVISION
→
SYSTEM
→
OPERATION
→
SIM. MEASURAND (
→
OFF)
Transducer_Error = No Error
OUT. SUBSTATUS = Non specific
No.
Error messages:
FOUNDATION Fieldbus
(FF)*
(local display)
Transducer Block
error messages
"Diagnosis" Transducer
Block
Analog Input Function Block
error messages
Reason for error/rectification
(spare parts
→
ä
71)