Liquiline System CA80HA
Technical data
Hauser
197
14.3.2 SD card
• Industrial Flash Drive, 1 GB
• Order number: 71110815
15
Technical data
15.1 Input
15.1.1 Measured variables
CaCO
3
[mg/l, µg/l, ppm, ppb]
15.1.2 Measuring range
0 to 80 mg/l (ppm) CaCO
3
15.1.3 Types of input
• 1 or 2 measuring channels (analyzer main parameter)
• 1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)
• Analog current inputs (optional)
• Binary inputs (optional)
15.1.4 Input signal
Depending on version
2 x 0/4 to 20 mA (optional), passive, potentially isolated
15.1.5 Current input, passive
Span
> 0 to 20 mA
Signal characteristic
Linear
Internal resistance
Non-linear
Test voltage
500 V
15.1.6 Cable specification (for optional sensors with Memosens technology)
Cable type
Memosens data cable CYK10 or sensor fixed cable, each with cable end sleeves or M12 round-
pin connector (optional)
Cable length
Max. 100 m (330 ft)