purpose, the instrument splits the bipolar manipulated variable y into two unipolar
manipulated variables, y+ and y
‐
.
Only the two unipolar manipulated variable parts are available for selection for outputting to
modulated relays. If outputting the values via a current output, you also have the possibility of
outputting the bipolar manipulated variable y to one current output only (split range).
10.7.2 Alarm relay and optional relays
The basic version of the device always has one alarm relay. Additional relays are also available
depending on the version of the device.
The following functions can be output via a relay:
• Limit switch status
• Controller manipulated variable for controlling an actuator
• Diagnostic messages
• Status of a cleaning function in order to control a pump or a valve
You can assign a relay to multiple inputs in order to clean several sensors with just one
cleaning unit, for example.
Menu/Setup/Outputs/Alarm relay or relay at channel no.
Function
Options
Info
Function
Selection
• Off
• Limit switch
• Controller
• Diagnostics
• Cleaning
(sensor)
• Formula
(sensor)
• Analyzer
Factory setting
• Alarm relays:
Diagnostics
• Other relays:
Off
The following functions depend on the option
selected.
These versions are illustrated individually in
the following sections to provide a clearer
understanding of the options.
Function
=
Off
Switches off the relay function and means no
further settings are required.