Liquiline System CA80HA
Diagnostics and troubleshooting
Hauser
155
No.
Message
Factory settings
Tests or remedial action
S
D
F
261
Electronics module
F
On
On
Electronics module defective
1.
Replace the module.
2.
Contact the Service Department.
262
Module connection
F
On
On
Electronics module not communicating
1.
Check the module , replace it if necessary.
2.
Contact the Service Department.
263
Incomp. detected
F
On
On
Wrong kind of electronics module
1.
Replace the module.
2.
Contact the Service Department.
284
Firmware update
M
On
Off
Update completed successfully
285
Update error
F
On
On
Firmware update failed
1.
Repeat.
2.
SD card error → use another card.
3.
Incorrect firmware → repeat with suitable
firmware.
4.
Contact the Service Department.
302
Battery low
M
On
Off
Buffer battery of real time clock is low
The date and time are lost if the power is
interrupted.
‣
Contact the Service Department (battery
replacement).
304
Module data
F
On
On
At least 1 module has incorrect configuration data
1.
Check the system information.
2.
Contact the Service Department.
305
Power consumption
F
On
On
Total power consumption too high
1.
Check installation.
2.
Remove sensors/modules.
306
Software error
F
On
On
Internal firmware error
‣
Contact the Service Department.
335
Fan defective
F
On
On
Fan defective
1.
Replace the fan.
2.
Contact the Service Department.