8. Route the cable in the housing in such a way that the
exposed
cable shield fits into one
of the cable clamps and the cable cores can be easily routed as far as the plug-in
terminals.
9. Open the cable clamp and clamp the cable in place. Then tighten the screw of the cable
clamp.
10. Connect the cable or cables (depending on the version) to the following plug-in
terminals:
SP2
N
L
PE
86
85
97
98
SP1
N
L
PE
86
85
97
98
BK
BK
GN/YE
GY
PK
GN
YE
1
2
1
2
A0028924
26
Connection scheme for Liquiline System CAT820 / 860
1
Hose heater 100 to 120 V/200 to 240 V AC (optional)
2
Connections for Memosens and communication with analyzer (optional)
SP1 Sample inlet
SP2 2nd sample inlet (optional)
11. Secure the protective cover after connecting. Make sure that no cables or hoses are
jammed.
12. Use the 6 screws to secure the carrier board after connecting.
6.4
Connecting the sensors and additional modules
6.4.1
Overview of connection compartment in the controller housing
The controller has a separate connection compartment.
‣
6 x Ph2
Release the 6 screws on the electronics compartment cover using a Phillips-head
screwdriver and fold out the cover towards the front.