Operation
Liquiline System CA80HA
100
Hauser
Menu/Setup/Inputs/Binary input x:y
Function
Options
Info
Signal type
=
Static signal
Signal level
Selection
• Low
• High
Factory setting
High
Specifies which input signal levels should
activate, for example, measuring range
switching or a cleaning.
Low
Input signals between 0 and 5 V DC
High
Input signals between 11 and 30 V DC
Signal type
=
PFM
Max. frequency
100.00 to 1000.00 Hz
Factory setting
1000.00 Hz
Maximum frequency of the PFM input signal
Is to equal the maximum possible upper limit
of the measuring range.
If the value selected is too small, higher
frequencies will not be detected. If the value is
too large, on the other hand, the resolution for
small frequencies will be relatively inexact.
Meas. value format
Selection
• #
• #.#
• #.##
• #.###
Factory setting
#.##
‣
Specify the number of decimal places.