Operation
Liquiline System CA80HA
82
Hauser
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Limits dispensers
Function
Options
Info
Alarm limits
Diag. code 732
Remaining operating time
Selection
1 to 60 d
Factory setting
7 d
An alarm limit can be configured for every
dispenser:
• Dispenser 2
• Dispenser 3
• Dispenser 7
Bottles
You can monitor the levels of all the bottles.
1. Switch on monitoring.
2. Enter the starting volume for every bottle.
3. Specify alarm and warning limits for the residual volumes.
4. Specify alarm and warning limits for the remaining time until the bottle refills are
consumed.
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Bottles
Function
Options
Info
Control
Selection
• Off
• On
Factory setting
Off
On
After the monitoring function is switched on,
you can find 2 additional menu items in:
Menu
/
Operation
/
Maintenance
/
Change
bottle
• Remaining operating time
• Bottle filling levels
Bottle filling levels
Start volume
Cleaner C
100 to 1000 ml
Factory setting
500 ml
Reagent RK
100 to 1000 ml
Factory setting
1000 ml
Standard S1
100 to 1000 ml
Factory setting
1000 ml
Warning limits
Diag. code 726
Warning limits are intended as an early
warning for maintenance tasks that will need
to be performed soon.
1.
Procure spare parts in time.
2.
Schedule maintenance tasks.