Liquiline System CA80HA
Operation
Hauser
85
10.3.2 Measurement
Menu/Setup/Analyzer/Measurement
Function
Options
Info
Start condition
Selection
• Immediate
• Date/time
• Continuous
Factory setting
Immediate
Immediate
: The analyzer immediately starts
with the measurement cycle once the system
changes to the automatic mode.
Date/time
: The analyzer starts with the
measurement cycle at the set date/time.
Continuous
: The analyzer measures
continuously, without interruption, between
the measurements.
Date
01.01.1970 to 07.02.2106
Functions only available for:
Start condition
=
Date/time
Time
00:00:00 to 23:59:59
Measuring interval
Factory setting
0:10
Function only available for:
Start condition
=
Immediate
or
Date/time
For configuring the time interval of the
measurement
Sequence of measurements
Selection
• SP1
• SP2
• Pause
Defines the sequence of channels in automatic
mode.
Is displayed only in the case of a two-channel
device.
•
SP1
: Indicate the number of consecutive
measurements for channel
SP1
•
SP2
: Indicate the number of consecutive
measurements for channel
SP2
•
Pause
: If the
Pause
channel is selected, the
analyzer does not perform the
measurement.
You can add, delete or save rows in the table
using the
INSERT
,
DEL
and
SAVE
soft keys.
Signal delay
0 to 600 s
Factory setting
0 s
Measurement active
: You can delay the
process of outputting the signal during an
active measurement by the defined time.
Measuring is paused for the duration of the
signal delay time.