Liquiline System CA80HA
Technical data
Hauser
207
15.7.8 Reagent requirement
• 75 μl (0.0025 fl.oz) per measurement
• Given a measuring interval of 10 min, a reagent lasts for approx. 90 days
15.7.9 Standard requirement
• 27 ml (0.91 fl.oz)per calibration
• Given a calibration interval of 72 h approx. 270 ml (9.13 fl.oz) per month
15.7.10 Calibration interval
1 h to 90 days, depending on the application and ambient conditions
15.7.11 Maintenance interval
Every 3 to 6 months, depending on the application
15.7.12 Maintenance effort
• Weekly: visual inspection
• Quarterly: 1 hour
15.8 Environment
15.8.1 Ambient temperature range
+5 to +40 °C (41 to 104 °F)
15.8.2 Storage temperature
–20 to 60 °C (–4 to 140 °F)
15.8.3 Humidity
10 to 95 %, non-condensing
15.8.4 Degree of protection
IP55 (cabinet, analyzer stand), TYPE 3R (cabinet, analyzer stand)
15.8.5 Electromagnetic compatibility
Interference emission and interference immunity as per EN 61326-1:2013, Class A for
Industry
15.8.6 Electrical safety
According to EN/IEC 61010-1:2010, Class I equipment
Low voltage: overvoltage category II
For installations up to 2000 m (6500 ft) above MSL
6)
Sufficient mains quality is required to operate the product as designated.