Operation
Liquiline System CA80HA
98
Hauser
Menu/Setup/Outputs/Binary output x:y
Function
Options
Info
Binary output
Selection
On
Signal type
Selection
Static signal
Function
Selection
Analyzer
Assignments
Function
=
Analyzer
Selection
Measurement active SP1
Here select which binary outputs output the
system status for a measurement currently
running.
In the case of two-channel devices,
Measurement active SP2
is also displayed
1)
x:y = slot no. : input number
9. Press
OK
to confirm.
The binary inputs and binary outputs are configured.
10. Switch to the automatic mode: Press the
MODE
soft key and select
Continue automatic
mode
or
Start automatic mode
.
The display shows
Current mode
-
Automatic
.
Activating an external pump before every measurement
If you convey the sample to the analyzer using an external pump or an external sample
preparation system, you can use the
Signal for sample request
to switch on the external
devices for a limited time only when the analyzer requires sample. The signal becomes active
before every measurement, calibration and cleaning. You can configure how long the signal
should be active for. The start of the actual analyzer activity is delayed by this time.
1. You must select
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Signal for sample request
.
2. Under
Lead time
, set how long an action that requires sample (measurement,
calibration or cleaning) is delayed.
3. Under
Duration SP%C
, set how long the signal should remain active. The duration can
be longer than the lead time. The maximum possible value equals the lead time plus the
duration of the measurement.
4. Configure the binary outputs as follows: