Liquiline System CA80HA
Operation
Hauser
81
Signals
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
Signal for sample request
Lead time SPX
Selection
0.00 to 30.00 (
mm:ss
)
Factory setting
0.00 (
mm:ss
)
In the automatic mode, each measurement
only starts once the lead time has elapsed. The
request sample signal is active from the start of
the lead time until the end of the specified
signal duration time.
Duration SPX
Selection
0:00 to 60:00 (
mm:ss
)
Factory setting
3:00 (
mm:ss
)
You can set the duration for which the signal is
active.
Signal for process access
SP1
Selection
• Always accessible
•
Binary input
x:y
Factory setting
Always accessible
Always accessible
: The input signal level at
the binary inputs does not affect actions that
require sample (measurement, calibration,
cleaning).
Binary input
x:y: When the signal is active at
the selected input, the analyzer starts actions
where sample is required. Otherwise the
analyzer defers or skips the action.
In the case of two-channel devices
SP2
is also
displayed.
1)
X is context-dependent, is automatically generated by the software and stands for the name of the sample
preparation, e.g. "1" or "2".
Diagnostic settings
Limit values for dispensers
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Limits dispensers
Function
Options
Info
Control
Selection
• Off
• On
Factory setting
On
Warning limits
Diag. code 733
Remaining operating time
Selection
1 to 90 (d)
Factory setting
28 d
A warning limit can be configured for every
dispenser:
• Dispenser 2
• Dispenser 3
• Dispenser 7