15.6.4 Cable entries
• 4 x bores for M16, G3/8, NPT3/8", Memosens connection
• 4 x bores for M20, G1/2, NPT1/2"
15.6.5 Cable specification
Cable gland
Permitted cable diameter
M16x1.5 mm
4 to 8 mm (0.16 to 0.32")
M12x1.5 mm (for order version M12 socket for Memosens sensors)
2 to 5 mm (0.08 to 0.20")
M20x1.5 mm
6 to 12 mm (0.24 to 0.48")
NPT³/₈"
4 to 8 mm (0.16 to 0.32")
G³/₈
4 to 8 mm (0.16 to 0.32")
NPT½"
6 to 12 mm (0.24 to 0.48")
G½
7 to 12 mm (0.28 to 0.48")
Cable glands mounted at the factory are tightened with 2 Nm.
15.6.6 Sensor connection (optional)
Sensors with Memosens protocol
Sensor types
Sensor cable
Sensors
Digital sensors
without
additional internal power
supply
With plug-in
connection and
inductive signal
transmission
• pH sensors
• ORP sensors
• Combined sensors
• Oxygen sensors (amperometric and optical)
• Conductivity sensors with conductive measurement of
conductivity
• Chlorine sensors (disinfection)
Fixed cable
Conductivity sensors with inductive measurement of
conductivity
Digital sensors with additional
internal power supply
Fixed cable
• Turbidity sensors
• Sensors for interface measurement
• Sensors for measuring the spectral absorption coefficient
(SAC)
• Nitrate sensors
• Optical oxygen sensors
• Ion-sensitive sensors