Operation
Liquiline System CA80HA
80
Hauser
10.3.1 Extended setup
Measured value
Menu/Setup/Analyzer/Extended setup/Measuring value
Function
Options
Info
Main value
Selection
• CaCO3
• Ca
Factory setting
CaCO3
Unit
Selection
• mg/l
• µg/l
• ppm
• ppb
Factory setting
mg/l
Main value format
Selection
• #
• #.#
• #.##
• Auto
Factory setting
#.#
Status for failed measurement
Selection
• Keep last value
• Status bad
Factory setting
Keep last value
Keep last value
Standard practice is to keep the measured
value of the last successful measurement until
the next successful measurement.
To detect an error during measurement, the
NAMUR status signal, the time stamp of the
measurement and the measured value status
must be evaluated.
Status bad
If a measurement cannot be performed due to
an error, the measured value status is set to
Bad
and the measured value to 0. The time
stamp is updated accordingly.
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
Device Range
Selection
The options available depend on the device
version ordered.
‣
Select the desired measuring range from
the list.