Operation
Liquiline System CA80HA
94
Hauser
2. In the software, configure a controller and as the source select the binary input
(e.g.
Binary input 1
) to which the proximity switch is connected. (
Menu
/
Additional
functions
/
Controllers
/
Controller 1
/
Controller Enable
=
Binary input 1
)
3.
Signal type
: For the selected input, select the factory setting (
Static signal
).
4. Connect the measured value of a flowmeter to the second input of the DIO module (e.g.
slot 6, port 2).
5.
Signal type
: for this input, select
PFM
. (
Menu
/
Inputs
/
Binary input
6:2
/
Signal type
=
PFM
)
6.
Input variable
: select the corresponding measured value (
Flow rate
).
You can now use the input you have just configured as the disturbance variable for
your controller
7.
Disturbance variable
: in the controller menu, select the binary input to which you have
connected the flow measured value. (
Menu
/
Additional functions
/
Controllers
/
Controller 1
/
Disturbance variable
/
Source of data
=
Binary input
6:2
and
Measured
value
=
PFM value
)
8. You can activate a dosing pump through PFM via a digital output of the DIO module.
Connect the pump to an output of the DIO module (e.g. slot 6, port 1) and select the
following settings in the menu:
Menu
/
Outputs
/
Binary output
6:1
/
Signal type
=
PFM
and
Source of data
=
Controller 1
.
Take the direction of action of your dosing into account. Select the correct parameter
(
Actuator type
=
U
or
Unipolar-
).
You must make additional settings in the controller menu to fully customize the control to suit
your process conditions.
1)
An activation code, Order No. 71211288, is necessary for the "feedforward control" function.