Maintenance
Liquiline System CA80HA
182
Hauser
5. Select
Confirm bottles inserted
.
Recommendation: Afterwards, perform a manual calibration with the standard solution
(
Determine calibration factor
).
Alternatively, the zero point and calibration factor can be determined in one step
(
Determine zero point and calibr. factor
).
12.2.6 Replacing the hoses
You require the following parts:
• From the CAV800 maintenance kit:
• NORPRENE hose, ID 1.6 mm
• C
‐
Flex hose, ID 3.2 mm
• C
‐
Flex hose, ID 6.4 mm
• Hose connector
• 1 pair of gloves that are resistant to the reagents used
1. Rinse the system. → 178
There should not be any more sample left in the sample collecting vessel.
2. Remove the bottles and the bottle tray.
3. Remove the cover of the carrier board.
4. Replace the hoses with hoses of the same diameter and length. Attach hose labels to the
new hoses.
5. Route drain hose D6
behind
dispenser 7.
6. Secure the cover and place the bottle tray back into the housing.
7. Connect the bottles to the appropriate covers and hoses.
8. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle
selection
.
9. Select all the bottles and press
OK
to confirm.
10. Select
Confirm bottles inserted
.
11. Press the
MODE
soft key and select
Continue automatic mode
or
Start automatic
mode
.
12.2.7 Replacing filter mats
You require the following parts:
Filter mats (part of the CAV800 maintenance kit)