Operation
Liquiline System CA80HA
116
Hauser
3. Define how the device should behave in the hold state (for sensors). (Configuration
options of
Source of data
,
Measured value
and
Hold behavior
Please note that if you select
Hold behavior
=
Freeze
, the system not only flags the status
but also actually "freezes" the measured value.
Further information on "Modbus" can be found in the Guideline for Communication via
Modbus, SD01189C
10.7.5 EtherNet/IP
Specify which process values should be output via EtherNet/IP communication.
You can define a maximum of 16 analog device variables (AI).
1. Define the data source.
You can choose from the analyzer as well as sensor inputs and controllers.
2. Select the measured value to be output.
3. Define how the device should behave in the hold state (for sensors). (Configuration
options of
Source of data
,
Measured value
and
Hold behavior
4. In the case of controllers, also specify the type of manipulated variable.
Please note that if you select
Hold behavior
=
Freeze
, the system not only flags the status
but also actually "freezes" the measured value.
In addition
you can define 8 digital device variables (DI):
‣
Define the data source.
You can choose from relays, binary inputs and limit switches.
For further information on "EtherNet/IP", see the Guideline for Communication via
EtherNet/IP, SD01293C
10.8 Additional functions
10.8.1 Limit switch
There are different ways to configure a limit switch:
• Assigning a switch-on and switch-off point
• Assigning a switch-on and switch-off delay for a relay
• Setting an alarm threshold and also outputting an error message
• Starting a cleaning function (for sensors)