Liquiline System CA80HA
Operation
Hauser
107
Menu/Setup/Outputs/Current output x:y
Function
Options
Info
Hold behavior
(for sensors)
Selection
• Freeze last value
• Fixed value
• Ignore
Factory setting
Depends on the channel:output
Freeze last value
The device freezes the last current value.
Fixed value
You define a fixed current value that is output
at the output.
Ignore
A hold does not affect this current output.
Hold current
(for sensors)
Hold behavior
=
Fixed value
0.0 to 23.0 mA
Factory setting
22.0 mA
‣
Specify which current should be output at
this current output in the hold state.
1)
x:y = slot:output number
Measured value
depending on the
Source of data
Source of data
Measured value
Total hardness
Hardness
Selection
Main value
pH Glass
Selection
• Raw value mV
• pH
• Temperature
pH ISFET
ORP
Selection
• Temperature
• ORP mV
• ORP %
Oxygen (amp.)
Selection
• Temperature
• Partial pressure
• Concentration liquid
• Saturation
• Raw value nA
(only
Oxygen (amp.)
)
• Raw value μs
(only
Oxygen (opt.)
)
Oxygen (opt.)
Cond i
Selection
• Temperature
• Conductivity
• Resistance
(only
Cond c
)
• Concentration
(only
Cond i
and
Cond c 4-pol
)
Cond c
Disinfection
Selection
• Temperature
• Sensor current
• Concentration