Liquiline System CA80HA
Operation
Hauser
91
Menu/Setup/Sample preparation/Installation type = Inlet pipe sampling/Sample preparation 1 ... 2
Function
Options
Info
Filter backflush
Selection
• On
• Off
Factory setting
On
Cleaning duration
Selection
10 to 60 s
Factory setting
00:30
Cleaning duration of the automatic water or compressed air
cleaning system
Cleaning discard time
Selection
0 to 1800 s
Factory setting
180 s
Discard time once cleaning is complete. If water is used as the
flushing medium, this must be replaced by a fresh sample
before starting the next measurement.
Lead time
0:00 to 30:00 min:ss
Factory setting
25:00 min
Duration
0:00 to 60:00 min:ss
Factory setting
25:00 min
Timeout no sample
3:00 to 30:00 min:ss
Reset settings
Resets all settings specific to sample preparation. All other
settings are retained.
10.5 Current inputs
The input can be used as the data source for limit switches and logbooks, for example.
Furthermore, external values can be made available as set points for controllers.
Menu/Setup/Inputs/Current input x:y
Function
Options
Info
Mode
Selection
• Off
• 0 - 20mA
• 4 - 20mA
Factory setting
4 - 20mA
‣
Select the same current range as in the
data source (connected device).
Input variable
Selection
• Parameter
• Current
Factory setting
Current
‣
Select the input variable.