7.4.3 Advanced Calibration
A sequence is executed step (action) by step
until
•
it reaches an action
END-OF-PRGM,
or
•
"
Action30
" has been executed, after which
the sequence quits automatically.
Example
The channels of an analyzer are to be cali-
brated in the following order:
All:
(Action1) Zero calibration
Channel 1:
(Action2) Span gas calibration
Channel2:
(Action3) Blowback
(Action4) Span gas calibration
Channel3:
(Action5) Span gas calibration range1
(Action6) Span gas calibration range3
... (Action7) End of program sequence
Program Sequence 1of8
Action1:
Zero-Cal
Node1:
All
Action2:
SpanCal
Node2:
Ch1
Action3:
Blowback
Node3:
Ch2
Action4:
SpanCal
Node4:
Ch2
Program Sequence 2of8
Action5:
Rg1SpanCal
Node5:
Ch3
Action6:
Rg3SpanCal
Node6:
Ch3
Action7:
END-OF-PGRM
Node7:
All
Action8:
END-OF-PGRM
Node8:
All
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......