Emerson Process Management GmbH & Co. OHG
9-47
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
9
Modbus Commands
9.4 List of Parameters and Registers - Sorted by Daniel Registers
Address
Tag Name
Data
Type
Client
Access
Description
Daniel Modicon
8043
5085…
5086
Status.NamurStates.MaintRequAl-
armComp5
DWord
RO
Namur MaintRequ alarms that are component5
related
8044
5087…
5088
Status.NamurStates.FctCheckA-
larmComp5
DWord
RO
Namur FctCheck alarms that are component1
related
8181
5361…
5362
Status.IO.ValveState
DWord
RO
Valve state bitfield (b0=V1, b1=V2, ....b19=V20)
9001
6001…
6002
PV1
Float
RO
Primary Variable 1
9002
6003…
6004
PV2
Float
RO
Primary Variable 2
9003
6005…
6006
PV3
Float
RO
Primary Variable 3
9004
6007…
6008
PV4
Float
RO
Primary Variable 4
9005
6009…
6010
PV5
Float
RO
Primary Variable 5
9006
6011…
6012
Service.Measurement.Simulation.
PVARawValue1
Float
RO
value for rawPVA of comp1
9006
6011…
6012
Status.Diagnostics.Raw-
QuotConce1
Float
RO
raw ADC quotient of component1
9007
6013…
6014
Service.Measurement.Simulation.
PVARawValue2
Float
RO
value for rawPVA of comp2
9007
6013…
6014
Status.Diagnostics.Raw-
QuotConce2
Float
RO
raw ADC quotient of component2
9008
6015…
6016
Service.Measurement.Simulation.
PVARawValue3
Float
RO
value for rawPVA of comp3
9008
6015…
6016
Status.Diagnostics.Raw-
QuotConce3
Float
RO
raw ADC quotient of component3
9009
6017…
6018
Service.Measurement.Simulation.
PVARawValue4
Float
RO
value for rawPVA of comp4
9009
6017…
6018
Status.Diagnostics.Raw-
QuotConce4
Float
RO
raw ADC quotient of component4
9010
6019…
6020
Service.Measurement.Simulation.
PVARawValue5
Float
RO
value for rawPVA of comp5
9010
6019…
6020
Status.Diagnostics.Raw-
QuotConce5
Float
RO
raw ADC quotient of component4
9011
6021…
6022
Status.Diagnostics.RawMeas-
Conce1
Float
RO
raw ADC of measure-side component1
9012
6023…
6024
Status.Diagnostics.RawRefConce1 Float
RO
raw ADC of reference side component1
9013
6025…
6026
Status.Diagnostics.RawMeas-
Conce2
Float
RO
raw ADC of measure-side component2
9014
6027…
6028
Status.Diagnostics.RawRefConce2 Float
RO
raw ADC of reference side component2
9015
6029…
6030
Status.Diagnostics.RawMeas-
Conce3
Float
RO
raw ADC of measure-side component3
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......