Emerson Process Management GmbH & Co. OHG
7-43
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
7
Maintenance & Procedures
7.4.7 Restoring a Calibration
Any menu, from where a calibration can be
started, shows a line
"Cancel!"
(see example
to the left).
Press
enter
in such a line to cancel any on-
going calibration. While canceling, a 'Function
executing' message appears.
7.4 Calibration Procedures
7.4.8 Cancelling an Ongoing Calibration
Ch1
Zero calibration
Cancel!
Start!
Zero gas
0.000 ppm
Concentration
0.000 ppm
Flow
0.00 l/min
Status..
Results..
Restore!
In case a wrong configuration was detected
after a calibration was carried out (e.g. wrong
gas connected), there is an option to restore
the last calibration data:
Any menu, from where a channel specific
calibration can be started, shows a line "
Re-
store!
" (see example to the left).
Press
enter
in such a line to restore the last
calibration data for the selected channel and
type of calibration (zero/span). While restore
is processing, a 'Function executing' message
appears.
Ch1
Zero calibration
Cancel!
Start!
Zero gas
0.000 ppm
Concentration
0.000 ppm
Flow
0.00 l/min
Status..
Results..
Restore!
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......