Emerson Process Management GmbH & Co. OHG
6-64
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Signal:
Comp1
OutRange:
0-20 mA
Low Scale:
0.00
Max Scale:
100.00
AutoScale:
Yes
FailMode:
Track
0/4 mA:
0.00
20 mA:
100.00
Hold:
No
"
Out(put) rang
e":
•
0-20 mA
(dead zero) generates a 20 mA
signal, if a concentration is measured at
the upper limit of the signal range. A 0 mA
signal is generated if the sample gas con-
centration equals the value specified with
"
LowScale
" .
•
4-20 mA
(life zero): A 4 mA signal is gene-
rated if the concentration equals the value
specified with "
LowScale
", thus enabling to
detect e.g. a broken cable.
"
Fail mode
" selects the output´s behaviour un
-
der failure conditions, considering or not, the
NAMUR recommendation NE 43. NE 43 defines
output signals enabling to detect different types
of failures/status (see table 6-1): The related
information is transmitted as a current signal,
but outside the (0)4-20 mA measurement signal
range.
Available options:
Track
: NE 43 not considered.
HIGH + 10%
: NE 43 failure signal level:
"above".
LOW - 10%
: NE 43 failure signal level:
"below".
6.2.3 Setup Menu
Enter a concentration, to equal the low output
limit (0 or 4 mA)
Enter a concentration, to equal the high output
limit (20 mA)
Enable (
Yes
)
or disable (
No
) output autosca-
ling.
Note!
Factory settings are OutRange: 4-20 mA and
FailMode: LOW - 10%, if not changed at time
of order.
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......