Emerson Process Management GmbH & Co. OHG
6-76
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Adjust Conversion Input1
Mode:
0..20mA
ADC
O
utput
8388608
Input Normalized to 0...1
Coeff a0:
1
Coeff a1:
2
Coeff a2:
3
Calc. Input Value
1013
Status
Good
Select the type of signal, provided by the
external source.
Available options:
0..10V, 0..1V, 0..20mA
: voltage or current
signal is applied
Off
: Input is switched off
Test
: the instrument internally generates a
step-like signal for testing purposes.
Enter the coefficients for your input signal
here, considering
- it is for a cubic equation max.
- it is normalized to 0...1 (see note in menu)
Shows the internal ADC output (informative)
Shows the calculated input value, based on
the coefficients a0..a2, for comparison with
the real input value. If both do not match,
correct the coefficients.
Input signal status.
Possible values:
Absent, Failure, Good, Simulated
(e.g.
when in test mode)
6.2.3 Setup Menu
6.2.3.4.5.1.1
Adjust Conversion
Setup..
In-/Outputs..
Analog Inputs..
Analog Input
1 ... 2
Adjust Conversion..
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......