Emerson Process Management GmbH & Co. OHG
5-16
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Multi-channel unit:
Select the component to be set in the gas
component selection menu.
Once the display settings have been chek
-
ked, press
left
to return to SETUP and
open
CALIBRATION to check the calibration
settings.
In
CALIBRATION - GASES, enter the values
for zero and span gas:
• See gas supplier´s certificate for correct
values.
• Values must be correctly set for results to
be accurate.
•
Multi-channel units: the values for each
channel must be entered separately.
5.7.3 Calibration Setup
5.7 Checking the Settings
Press
left
to return to
CALIBRATION
, and
enter
TOLERANCES
.
By default the 'Deviation Tolerance check
'
option is set inactive (
Off
).
With "
DeviatToler"
set to
On
,
• during calibration the analyzer checks
whether the values set for zero and span
gas conform to the concentration of the gas
currently being supplied.
•
If the concentrations differ more than the
percentage of range entered in the follo-
wing lines, the calibration is aborted.This
prevents calibration from being performed
when the incorrect gas is supplied (e.g.
span gas calibration using zero gas), which
would result in an incorrectly configured
unit.
Note!
For more detailed information on calibration
procedures,
Setup..
Calibration..
Ch1
Calibration
Gases..
Tolerances..
Procedure..
Valve Assignment..
Program Sequence..
Interval Time..
Deviations..
Gases
Zero Gas:
0.000 ppm
Span Gas:
50000.000 ppm
Range Gases..
Current Range:
Range 1
Ch2
Tolerances
DeviatToler:
Off
Zero Limit:
20.0 %
Span Limit:
20.0 %
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......