Signal:
Comp1
OutRange:
0-20 mA
Low Scale:
0.00
Max Scale:
100.00
AutoScale:
Yes
FailMode:
Track
0/4 mA:
0.00
20 mA:
100.00
Hold:
No
5.7 Checking the Settings
"
0/4 mA
" enables to finetune the analog
output: Set "
Signal
" to
0 mA
and, while
measuring the output current, adjust it to the
expected value.
Accepted range:
-10,000
…
+10,000
"
20 mA
" enables to finetune the analog
output: Set "
Signal
" to
20 mA
and while
measuring the output current, adjust it to the
expected value.
Accepted range:
-10,000
…
+10,000
"
Hold
" selects the output´s behaviour during
calibrations.
If set to
Yes
,
• the analog output is fixed to the last measu
-
red value;
•
concentration alarms, which may otherwise
be triggered by the concentrations of the
calibration gases, are supressed.
If set to
No
,
•
the analog output signal always corre-
sponds to the actual measured value during
calibration; this may trigger alarms if limits
are exceeded.
Note!
This behaviour may be undesireable if e.g.
the unit is connected to a data acquisition
system.
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......