Emerson Process Management GmbH & Co. OHG
3-16
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
3.4 Trace Moisture Measurement
3.4.1 Special Operating Conditions
The sensor is completely calibrated with all
calibration data stored in its flash memory and
does not require recalibration:
• If the sensor is included into a calibration
procedure, it might end up with a wrong
calibration and unusable sensor. Therefore
the analyzer's trace moisture measurement
channel is configured to be excluded from
autocalibration procedures, by default cali
-
brating all channels. This exclusion is done
by factory setup and cannot be changed.
•
For proper measurement results we re-
commend to exchange the sensor regularly
after 12 months of operation.
How does dew point measurement connect
with trace moisture measurement?
The lower the dew point of a gas, the less is
the content of moisture within that gas.
dp / °C Water / ppm
-100
0.025
-98
0.038
-96
0.057
-94
0.084
-92
0.123
-90
0.179
-88
0.258
-86
0.368
-84
0.520
-82
0.729
-80
1.01
-78
1.40
-76
1.91
-74
2.59
-72
3.49
-70
4.68
-68
6.22
-66
8.22
-64
10.8
-62
14.1
-60
18.3
-58
23.5
-56
30.2
-54
38.5
-52
48.9
-50
61.8
-48
77.6
-46
97.1
dp / °C Water / ppm
-44
121
-42
150
-40
185
-38
228
-36
279
-34
340
-32
413
-30
501
-28
604
-26
726
-24
870
-22
1039
-20
1237
-18
1468
-16
1737
-14
2048
-12
2409
-10
2826
-8
3306
-6
3856
-4
4487
-2
5208
0
6030
2
6964
4
8025
6
9226
8
10 583
10
12 113
Tab. 3-6:
Dew Points and Water Content (at 1013 hPa)
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......