Troubleshooting
Message
Status level
Explanation
Recommended Actions
Factory file read error
Maintenance
Reading the factory configu
-
ration file is not possible
•
Check if internal disk is pre-
sent / installed
•
Call Emerson service
Factory file write error
Maintenance
Writing the factory configura
-
tion file is not possible. This
message does not appear
during normal operation!
FATAL!! Configuration
data destroyed
Failure
The instrument is now uncon-
figured, because retrieving
the configuration data from
several sources failed.
FATAL: FRAM read/write
error
Maintenance
The instrument is now uncon-
figured, because retrieving the
configuration data from sever-
al sources failed.
Call Emerson service
Flash write count over limit
Maintenance
Write cycles to internal CPU
Flash Memory exceeded
number of 90,000
Limitation analog output 1
Off-spec
Concentration assigned to
the indicated analog output
is outside configured ranges:
Analog output is limited to
configured ranges
•
Use another measurement
range.
•
Extend analog output range
configuration if possible.
•
Run measurement inside its
given ranges.
Limitation analog output 2
Off-spec
Limitation analog output 3
Off-spec
Limitation analog output 4
Off-spec
Limitation analog output 5
Off-spec
PLC program error
Maintenance
While installing the PLC pro-
gram, a program error was
discovered
Check the PLC program for
syntax errors, wrong com-
mands or references
8.2.1 Analyzer Related Messages
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......