Emerson Process Management GmbH & Co. OHG
1-7
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
1.2 Configuration of
Gas Lines
1.2.6.5 Internal Temperature Sensors
In the same way as pressure variations, va-
rying temperatures influence the measuring
results: Higher temperature results in lower
gas density and thus in less molecules per
volume. To compensate temperature influ
-
ence, internal temperature sensors can be
installed to electronically compensate tempe-
rature variations (
specification).
1.2.6.6 Internal Barometric Pressure Sensor
Varying atmospheric pressure has an influ
-
ence also on the density of the gases applied
to the measuring system: Higher density cor-
relates with more molecules per volume and
thus influences the measuring results.
To compensate such influences. an internal
barometric pressure sensor can be installed.
It´s reading is used to electronically compen-
sate the atmospheric pressure variation (
measurement specification, page 3-21).
If such a sensor is installed in the unit, the
related menu shows the entry
Internal
(
Depending on the configuration of the unit or
the demands of the application, temperature
sensors can measure the unit’s internal tem-
perature or selected measurement channel
components.
If such sensors are installed in the unit, this
is indicated in the installed options
menu
(
1
T
echnical
Description
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......