Emerson Process Management GmbH & Co. OHG
7-30
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
7.4.3 Advanced Calibration
The analyzer immediately begins to calibrate,
showing the
CALIBRATION STATUS SUM
-
MARY screen.
"
Current action"
indicates, what currently is
carried out (
purging, zeroing, spanning,
ready
)
"
Action detail"
shows the channel currently
calibrated
"
Current duration
" gives the remaining time
for the current procedure
"
Prev. duration
" shows the time elapsed since
start of procedure
To see a detailled calibration status for a sin
-
gle channel, enter
CALIBRATION STATUS
SINGLE,
Calibration status summary
Calibration status single..
Current action
Ready
Action detail
Ch2
Current duration
0 s
Prev. duration
0 s
Current Step
0
Multi-channel unit:
Select the channel in SELECT COMPO-
NENT.
to open the status screen with enhanced
calibration information about the current
channel (indicated in the uppermost display
line), including
remaining calibration time, currently measu
-
red concentration, setup of zero & span gas
concentrations and currently calibrated range.
The procedure has finished when "
Applied
gas"
shows
Sample gas
, or "
Current action
"
in the previous screen says
Ready
.
Press
home
to return to the MEASUREMENT
SCREEN.
Ch1
Calibration status single
Cancel!
Calibr.status
Ready
Remaining time
0 s
Concentration
0.000 ppm
Zero gas
0.000 ppm
Span gas
5000.000 ppm
Current range
Range 1
Applied gas
Zero gas
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......