X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
7
Maintenance & Procedures
7.4.5 Unattended Automatic Calibration
All submenus opened by above menu lines
have the same content, exemplarily described
below:
Set to
Enabled
, to use interval times for the
selected calibration (here: Zero all)
Enter the interval time
Accepted range:
1 ... 10,000
h
In this submenu enter the start time for the
first calibration (s. below)
In this submenu enter the date and start
time for the first calibration after finishing this
setup.
Note!
Time format is 24h (1 pm = 13)
Press
left
to return to the previous menu, to
see a summary.
Start time & date for the next calibration,
based on the entered parameters (empty
until date & time have been entered in the
submenu)
Start time & date for the next calibration,
based on the entered parameters
Current date & time
...
Enable:
Enabled
Interval:
15 h
Start time..
Next:
--
Time
10/01/10 15:33
...
Enable:
Enabled
Interval:
15 h
Start time..
Next:
10/01/10 16:00
Time
10/01/10 15:33
Start...
Month:
1
Day:
10
Hour:
16
Minute:
0
Set!
Next
10/01/10 16:00
Note!
If the displayed current time is not correct,
update the system setup on
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......