Emerson Process Management GmbH & Co. OHG
8-6
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Message
Status level
Explanation
Recommended Actions
SCAL blowback
Function check
Device runs system
calibration’s blowback mode
• Wait until system calibration
procedure is finished
• Cancel system calibration
procedure
SCAL program sequence
Function check
Device runs system
calibration’s program se
-
quence
SCAL spanning
Function check
Currently a system
calibration’s spanning is on
-
going
SCAL zeroing
Function check
Currently a system
calibration’s zeroing is on
-
going
• Wait until system calibration
procedure is finished
• Cancel system calibration
procedure
SCAL zeroing & spanning
Function check
Currently a system calibrati
-
on (zeroing and spanning) is
ongoing
Sensor CRC-check
Failure
XPSV - CPU communication
failure
Call Emerson service, if mes-
sage shows up repeatedly
Sensor command buffer
overflow
Failure
Sensor failure
Failure
Sensor invalid message
length
Failure
SENSOR RESET
Maintenance
8.2.1 Analyzer Related Messages
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......