Emerson Process Management GmbH & Co. OHG
6-67
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
6
Software Menus
Cut Mode:
Standard
Low Cut:
3.80 mA
High Cut:
20.50 mA
6.2.3 Setup Menu
Note!
This second menu page appears only, if
"Fail-
Mode"
is set to other than Track! It enables
to specify the output´s behaviour in case the
measured value exceeds the range (
.
"Low Cut"
is output if the measured value
is below the lower range limit,
"High Cut"
is
output if it exceeds the upper range limit.
"
Cut Mode
" specifies if
Standard
values or
Config
urable values are used for output:
If
"Cut Mode"
is set to
Standard
, these lines
show the standard settings. For
"Low Cut"
they depend on the setting of
"OutRange"
:
OutRange
0-20 mA
4-20 mA
Low Cut
-0.20 mA
3.80 mA
High Cut
20.50 mA
If
"Cut Mode"
is set to
Config
, use these
lines to adjust the related output signals. The
accepted values for
"Low Cut"
again depend
on the setting of
"OutRange"
:
OutRange
0-20 mA
4-20 mA
Low Cut
-1.80…-0.01 mA 2.20…3.99 mA
High Cut
20.01…21.50 mA
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......