Emerson Process Management GmbH & Co. OHG
6-49
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
6
Software Menus
Ch1
Linearization
Linearizer:
Off
LinearMethod:
Splines
Calculate!
Setpoints (X)..
Actuals (Y)..
Linearizer Min
20.000 ppm
Linearizer Max
110.000 ppm
Status
Normal
6.2.3.3.3
Setup Linearization
6.2.3 Setup Menu
Specify the linearization methode to be
used.
Available options:
Splines
,
Polynom
.
Switch
On
or
Off
the linearizer.
First enter
SETPOINTS (X)
to enter the new
setpoint values (
When done, enter ACTUALS (Y) to enter the
corresponding actuals (
Select this line to let the analyzer calculate
the new linearization curve.
The linearization status.
Possible values:
Normal, Underflow, Over
-
flow, Undefined
Setup..
Measurement..
Linearization..
If the system is set up accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......