Problem
Possible causes
Recommended actions
Unusually high densi-
ty reading
•
Plugged or coated sensor tube
•
Incorrect density calibration factors
•
Incorrect temperature measurement
•
RTD problem
•
In high-frequency meters, this can indicate
erosion or corrosion
•
In low-frequency meters, this can indicate
tube fouling
•
Ensure that all of the calibration parame-
ters have been entered correctly. See the
sensor tag or the calibration sheet for your
meter.
•
Verify all of the characterization or calibra-
tion parameters. See the sensor tag or the
calibration sheet for your meter.
•
Purge the sensor tubes.
•
Check for coating in the flow tubes.
Unusually low density
reading
•
Two-phase flow
•
Ensure that all of the calibration parame-
ters have been entered correctly. See the
sensor tag or the calibration sheet for your
meter.
•
In low-frequency meters, this can indicate
erosion or corrosion
•
Check your process conditions against the
values reported by the device.
•
Check for two-phase flow.
•
Verify all of the characterization or calibra-
tion parameters. See the sensor tag or the
calibration sheet for your meter.
•
Check the wiring between the sensor and
the transmitter.
•
Check for tube erosion, especially if the
process fluid is abrasive.
12.5
Temperature measurement problems
Problem
Possible causes
Recommended actions
Temperature reading
significantly different
from process temper-
ature
•
RTD failure
•
Incorrect compensation factors
•
Line temperature in bypass does not
match temperature in main line
•
Check junction box for moisture or verdi-
gris.
•
Perform RTD resistance checks and check
for shorts to case (see
).
•
Verify that the temperature compensation
factors match the value on the sensor tag
or calibration sheet.
•
If Alert A004, A016, or A017 is active, per-
form the actions recommended for that
alert.
Troubleshooting
Configuration and Use Manual
173
Содержание Micro Motion 2500
Страница 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 20: ...Configuration and commissioning 12 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 124: ...Complete the configuration 116 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 130: ...Set up the Weights Measures application 122 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 132: ...Operations maintenance and troubleshooting 124 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 140: ...Transmitter operation 132 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 216: ...Using ProLink III with the transmitter 208 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 226: ...Default values and ranges 218 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Страница 237: ...NE 53 history Configuration and Use Manual 229 ...